Wang Cheng, Cheng Junxi. Capillary Ion Chromatographic Method in the Detection of Cations in Water[J]. Journal of Environmental Hygiene, 2012, 2(4): 187-188.
    Citation: Wang Cheng, Cheng Junxi. Capillary Ion Chromatographic Method in the Detection of Cations in Water[J]. Journal of Environmental Hygiene, 2012, 2(4): 187-188.

    Capillary Ion Chromatographic Method in the Detection of Cations in Water

    • Objectives Objective To explore the method of capillary ion chromatography for the detection ofcommon cations in water.
      Methods The characteristics of capillary ion chromatography in the determination ofpotassium, sodium, calcium and magnesium ion in water were studied.
      Results Under the optimal flow rate andoptimal concentration of leaching liquid methane sulfonic acid(MSA)at 0.018 mL/min and 28 mmol/L respectively, the peak area repeatability of potassium, sodium, calcium, magnesium ion in standard solution was all < 1%;and when the sample volume was 0.4 μL, the detection limit was < 3 μg/L.
      Conclusion The characteristic of the method was good in reproducibility, low in detection limit and high in sensitivity within the linear range incomparison with the traditional ion chromatography, and was also reagent saving and easier in operation.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return