甘平胜, 曾涛, 李晓晶, 刘苗, 杨荣, 李静, 潘心红. 吹扫捕集-GC-MS/MS法测定饮用水中2-甲基异莰醇和土臭素[J]. 环境卫生学杂志, 2023, 13(12): 923-927. DOI: 10.13421/j.cnki.hjwsxzz.2023.12.009
    引用本文: 甘平胜, 曾涛, 李晓晶, 刘苗, 杨荣, 李静, 潘心红. 吹扫捕集-GC-MS/MS法测定饮用水中2-甲基异莰醇和土臭素[J]. 环境卫生学杂志, 2023, 13(12): 923-927. DOI: 10.13421/j.cnki.hjwsxzz.2023.12.009
    GAN Ping-sheng, ZENG Tao, LI Xiao-jing, LIU Miao, YANG Rong, LI Jing, PAN Xin-hong. Determination of 2-methylisoborneol and geosmin in drinking water by purge and trap-GC-MS/MS[J]. Journal of Environmental Hygiene, 2023, 13(12): 923-927. DOI: 10.13421/j.cnki.hjwsxzz.2023.12.009
    Citation: GAN Ping-sheng, ZENG Tao, LI Xiao-jing, LIU Miao, YANG Rong, LI Jing, PAN Xin-hong. Determination of 2-methylisoborneol and geosmin in drinking water by purge and trap-GC-MS/MS[J]. Journal of Environmental Hygiene, 2023, 13(12): 923-927. DOI: 10.13421/j.cnki.hjwsxzz.2023.12.009

    吹扫捕集-GC-MS/MS法测定饮用水中2-甲基异莰醇和土臭素

    Determination of 2-methylisoborneol and geosmin in drinking water by purge and trap-GC-MS/MS

    • 摘要:
      目的 利用吹扫捕集-气相色谱串联质谱(GC-MS/MS)法测定生活饮用水中2-甲基异莰醇和土臭素的分析方法。
      方法 取5 mL水样,以高纯氮气会载气在40 mL/min流速下吹扫11 min后,热解吸进入GC-MS/MS进行分析,以2-异丁基-3-甲氧基吡嗪为内标进行定量。
      结果 在优化的实验条件下,2-甲基异莰醇和土臭素在5.0~100 ng/L浓度范围内线性较好,相关线性系数(R2)分别为0.999 6和0.999 8,方法检出限分别为0.13 ng/L和0.06 ng/L,水样加标回收率在92.3%~104.5%之间,相对标准偏差(RSD)为1.89%~4.07%。
      结论 该方法简便、快速、准确、灵敏度高,可用于测定饮用水中2-甲基异莰醇和土臭素。

       

      Abstract:
      Objective To develope a purge and trap-gas chromatography-triple quadrupole mass spectrometry(GC-MS/MS) method for determination of 2-methylisoborneol and geosmin in drinking water.
      Methods After 5 mL water sample was purged with high purity nitrogen carrier gas at a flow rate of 40 mL/min for 11 min, the residue is thermally desorbed and analyzed by GC-MS/MS. The 2-isobutyl-3-methoxypyrazine was used as the internal standard for quantification Result Under the optimized experimental conditions, the linearity of 2-methylisobornol and geosmin in the concentration range of 5.0 ng/L~100 ng/L was good, with a correlation coefficient of 0.999 6 and 0.999 8, respectively. The limits of detection (LODs, S/N=3) were 0.13 ng/L for 2-methylisoborneol and 0.06 ng/L for geosmin. The recoveries ranged from 92.3% to 104.5%, and the relative standard deviations ranged 1.89% to 4.07%.
      Conclusion The method is simple, rapid, accurate, and sensitive. It can be applied for the determination of 2-methylisobcamol and geosmin in drinking water.

       

    /

    返回文章
    返回